WB01343_3.gif (599 bytes)   Proper use of the Compound Microscope   WB01345_3.gif (616 bytes)
     Kohler Illumination & Staining-2
201 Syllabus

 

Locate the major parts of your microscope by referring to Illustrations below.

LeitzSScopeOverL-1-300.jpg (73303 bytes) LeitzSScopeOverL-2-400.jpg (99379 bytes)

We will be using Leitz microscopesLeitzSScopeSTransL300.jpg (94898 bytes)
in this class. The instructions for correct alignment of the condenser will be applicable to other microscopes with adjustable condensers.

Light is provided by a built in bulb which is reflected through the field iris diaphragm, the condenser, the condenser iris diaphragm, the specimen, the objective, the tube and the ocular.

There are various control knobs on the microscope which affect the light path. In addition, there are knobs for coarse and fine focus, as well as knobs to move the stage.

Focusing the Objectives

Locate the coarse and fine focusing knobs
on each side of your scope. Each knob does
coarse and fine focusing
. There is no
separate knob for fine focusing. You will
see how this works later. The rotation of
LeitzSScopeCondLab-1-300.jpg (74869 bytes)
this knob focuses the objective onto the specimen.

Mechanical Stage

The knobs which control the mechanical stage are on the right side of the microscope as it faces you.

Condenser

The condenser aligns and focuses light on the specimen. It has a long vertical knob
[pointing down] on its left side as you face the scope. Rotation of the small knob at its tip, raises and lowers the condenser to achieve focus.

Swinging Lens

Directly above this at a right angle to the
condenser adjusting knob you will find a
ZeisStScopeCondensLab-2-500.jpg (152223 bytes)
knob which controls a lens that can be swung in or out of place. This swinging lens is left out for low-power illumination [i.e. 4X], and swung into the light path for objectives of 10X or greater magnification. Failure to use this lens properly is the most common mistake that most people make. If you fail to use this properly you will not be able to see much, especially when we use thick
sections or whole mounts.

Centering Screws

In addition, there are two small knobs on the
front of the condenser, set at 45o which are
used to center it.

Condenser Iris Diaphragm

Finally, there is a lever which controls the
aperture of the Condenser Iris. This improves contrast (difference between light & dark)
especially at intermediate and high magnifications.

Do NOT use this to increase or decrease brightness!!!

Locate all of the controls for the condenser.

REMEMBER THIS !!!!!!!!!!!!!!!!!!!!!!!!!!

The Objective focuses on the Specimen
from
Above

The Condenser focuses light onto the
Specimen
from
Below.

Field Diaphragm

The light source is housed in the base of the microscope. It passes through the field iris diaphragm. The size of the field diaphragm is controlled by rotating a knurled ring which is concentric with it. The field diaphragm controls the area of illumination.

Image68.jpg (75741 bytes) ZeissSScopField Diaph300.jpg (60846 bytes)

Locate the field diaphragm and its knurled ring.
Page Top